Using the Adv Fctn Key
Advanced Function Descriptions
46
The sampler waits for 6850/6890 GC ready signal before beginning the
sequence. The headspace sampler’s Sample Equilibration Time is ignored;
the sequence is controlled by GC readiness.
CHECK GC READY YES -L 5890
The sequence aborts if the HP 5890 GC is not ready after the Vial Equib.
time has expired.
CHECK GC READY YES -H 6890
The sequence aborts if the 6850/6890 GC is not ready after the Vial Equib.
time has expired.
Note that “L” stands for Low and “H” stands for High.
If CHECK GC READY YES is active, then the sampler will check for the
presence of a ready signal from the GC just prior to injection.
If a GC ready signal is not found, the sampler will abort the run or sequence,
return all vials to the tray, and display the message GC NOT READY. If CHECK
GC READY NO is active, the headspace sampler will inject regardless of the
ready state of the GC.
7694 RESET
G1289/G1290 Adv Fctn, ., 1, Enter
Select 7694 Reset to force the sampler to abort whatever it is doing and re-run
its startup routine. See “Power on Diagnostics” on page 14 for more
information.
Note:
• The instrument remains Not Ready until all setpoints for the method are
reached.
• The last method in use before reset remains in effect.