Splitter Configurations 3
Installation and Operation Guide 37
Disconnect tubing from the splitter
Loosen and remove the internal nut from the splitter fitting. Usually the
tubing and ferrule will fall out of the fitting.
Occasionally the ferrule will stick in the fitting. If this happens, use a pointed
object like a pen or a paper clip and insert it in the ferrule release hole in the
side of the fitting (Figure 17). Press firmly. The ferrule will click when it
breaks free.
Figure 16 Tightening the connections
15°
Figure 17 Releasing a ferrule