Agilent InfinityLab LC Series Vialsamplers User Manual 117
3Using the Module
Install the Optional Sample Cooler/Sample Thermostat
Install the Sample Cooler/Sample Thermostat
Tools required p/n Description
8710-0899 Screwdriver Pozidrive Shaft
(for the Sample Cooler)
5182-3466 Torx screwdriver T10
(for the Sample Thermostat)
OR 5023-3089 Torx key set
Parts required # p/n Description
1 Sampler
1 G7167-60005 Sample Cooler
OR 1 G7167-60101 Sample Thermostat
1 Power cord
1 5067-6208 Condensate Drainage Kit
Preparations
Sampler is installed in the stack.
WARNING
Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat
and laboratory.
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m
3
(1 m
3
for every 8 g of R600a refrigerant
inside of the Sample Thermostat).
Ensure adequate ventilation: typical air exchange of 25 m
3
/h per m
2
of
laboratory floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do
not block the openings on the circumference of the Sample Thermostat.
WARNING
Flammable refrigerant used
When handling, installing and operating the Sample Thermostat, care
should be taken to avoid damage to the refrigerant tubing or any part of
the Sample Thermostat.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129C and is the answer not in the manual?

Agilent Technologies G7129C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals