Agilent InfinityLab LC Series Vialsamplers User Manual 249
8Maintenance
Exchange the Rotor Seal
9 Reconnect the capillaries and tubes to the valve ports.
The correct plumbing is shown below and on the safety
cover of the module.
10 Slide the waste tube into the waste holder in the leak
tray.
11 Close the doors. 12 Perform a System Pressure Test.
Waste
Column
Pump