EasyManuals Logo

Agilent Technologies G7129C User Manual

Agilent Technologies G7129C
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #264 background imageLoading...
Page #264 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 264
8Maintenance
Exchange the Wash Port Assembly
3 Verify that the needle in the uppermost position before
continuing.
4 Loosen the safety cover screw.
WARNING
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
5 Remove the safety cover.
WARNING
Sharp needle
Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
6 Move the wash port into the service position by slightly
lifting it up and then flapping to the left.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129C and is the answer not in the manual?

Agilent Technologies G7129C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals