WORKING WITH NORMAL-PHASE COLUMNS
Every normal-phase column is tested before shipment and shipped in a normal-
phase test eluent, with the exception of ZORBAX Eclipse XDB-CN and, InfinityLab
Poroshell 120 EC-CN which are shipped in reversed-phase solvents and
InfinityLab Poroshell 120 HILIC and ZORBAX HILIC Plus which are shipped in
HILIC solvents. ZORBAX carbohydrate columns are shipped in acetonitrile. Care
should be taken to make sure the column has been properly equilibrated prior
to use. This will ensure reproducibility from analysis to analysis and help prevent
retention time drifting.
Phase Recommended
pH Range
Maximum
Operating
Temperature
Suggested
Operating
Temperature
InfinityLab Poroshell
120 EC-CN
pH 2.0 to 8.0 60 °C 30 °C
ZORBAX Eclipse
XDB-CN
pH 1.0 to 9.0 60 °C 30 °C
ZORBAX CN pH 2.0 to 8.0 60 °C 30 °C
ZORBAX
Carbohydrate
pH 2.0 to 8.0 60 °C 30 °C
ZORBAX NH
2
pH 2.0 to 7.0 60 °C 30 °C
ZORBAX Rx-SIL,
ZORBAX SIL
pH 1.0 to 8.0 Limited only by the
temperature limits of
the mobile phase
Pursuit, Po-
laris, Metasil, Mono-
chrom, Microsorb
pH 2.0 to 8.0 60 °C 30 °C
InfinityLab Poroshell
120 HILIC, ZORBAX
HILIC Plus
pH 1.0 to 8.0 Limited only by the
temperature limits of
the mobile phase
Normal-Phase Column Operating Parameters:
pH and Temperature
Operating at extreme ends of pH and temperature ranges will have a significant impact
on column lifetime.