Operation Quick Reference Guide
Sampler Pre-run Checklist
Use this checklist to make sure the sampler is ready before you begin a run.
Sample vials are half full.
Cap is centered, no wrinkles, septum is flat.
Sample vial positions match the run parameters.
Tray quadrants snapped in place (with tray).
4.5 mL of fresh solvent in each solvent bottle.
Waste bottles are empty.
Two waste bottles (with tray).
There is enough solvent and waste capacity for your sample vials.
Syringe is new or clean.
Syringe design and size are correct.
Plunger is secure in plunger carrier loop.
Needle is aligned with septum retainer nut.
Syringe is rinsed with solvent.
GC inlet liner is clean and deactivated.
GC inlet liner is correct type for injection technique.
GC inlet septum type is correct and the septum is less than 200 injections old.
Correct septum nut is installed in GC inlet.
Correct nanoliter enable setting for required injection volume.
(GC or integrator control only)
Correct cool on-column inlet insert installed for your column/syringe.
Running a sample
To operate your 7683 Automatic Liquid Sampler:
1. Install a clean syringe. See Installing A Syringe on page 16.
2. Fill the solvent bottles and load them into the turret. See Preparing The Solvent And
Waste Bottles on page 13.
3. Load the waste bottles into the turret. See Preparing The Solvent And Waste Bottles on
page 13.
4. Load the sample vials into the turret or into the tray. See Preparing Sample Vials on
page 9.
5. Load (or program) the GC sequence. See your 6890 GC or Agilent ChemStation
documentation.
6. Run the GC sequence.
When the GC becomes ready, the 7683 Automatic Liquid Sampler begins the injections.