Operation Quick Reference Guide
Sampler Pre-run Checklist
Use this checklist to make sure the sampler is ready before you begin a run.
Sample vials are half full.
Cap is centered, no wrinkles, septum is flat.
Sample vial positions match the run parameters.
Tray quadrants snapped in place (with tray).
4.5 mL of fresh solvent in each solvent bottle.
Waste bottles are empty.
Two waste bottles (with tray).
There is enough solvent and waste capacity for your sample vials.
Syringe is new or clean.
Syringe design and size are correct.
Plunger is secure in plunger carrier loop.
Needle is aligned with septum retainer nut.
Syringe is rinsed with solvent.
GC inlet liner is clean and deactivated.
GC inlet liner is correct type for injection technique.
GC inlet septum type is correct and the septum is less than 200 injections old.
Correct septum nut is installed in GC inlet.
Correct nanoliter enable setting for required injection volume.
(GC or integrator control only)
Correct cool on-column inlet insert installed for your column/syringe.
Running a sample
To operate your 7683 Automatic Liquid Sampler:
1. Install a clean syringe. See Installing A Syringe on page 16.
2. Fill the solvent bottles and load them into the turret. See Preparing The Solvent And
Waste Bottles on page 13.
3. Load the waste bottles into the turret. See Preparing The Solvent And Waste Bottles on
page 13.
4. Load the sample vials into the turret or into the tray. See Preparing Sample Vials on
page 9.
5. Load (or program) the GC sequence. See your 6890 GC or Agilent ChemStation
documentation.
6. Run the GC sequence.
When the GC becomes ready, the 7683 Automatic Liquid Sampler begins the injections.

Other manuals for Agilent Technologies 7683

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7683 and is the answer not in the manual?

Agilent Technologies 7683 Specifications

General IconGeneral
BrandAgilent Technologies
Model7683
CategoryLaboratory Equipment
LanguageEnglish

Summary

Sampler Pre-run Checklist

Safety Information

Safety Symbols

Explains symbols used in the manual for warnings and cautions.

Electromagnetic Compatibility

Details CISPR 11 compliance and user guidelines for interference.

Sound Emission Certification for Federal Republic of Germany

Provides sound pressure level information relevant to German regulations.

About your 7683 Automatic Liquid Sampler

Capabilities

Injection Volume

Pre-run Checklist

Running A Sample

Interrupting A Run Or Sequence

Sampler Response To Interruptions

Explains how the sampler reacts to different types of interruptions.

Restarting An Interrupted Sequence

Running A Priority Sample

Preparing Sample Vials

Selecting Sample Vials

Discusses vial types, materials, and dimensions for compatibility.

Labeling Sample Vials

Sample Vial Septa

Describes septa types, properties, and suitability for samples.

Filling Sample Vials

Vial Placement Without 100-vial Tray

Details vial placement in standard or 8-position turrets.

Using The Optional 100-vial Tray

Preparing The Solvent And Waste Bottles

Using Two Injectors

With A Tray

Setup and injection cycle with a tray.

Without A Tray

Injection process and sample handling without a tray.

Syringes

Inspecting A Syringe

Checks for needle straightness, surface roughness, and plunger movement.

Installing A Syringe

Removing A Syringe

Cool On-column Injection

Controlling Sample Vial Temperatures

Connect the tray quadrants

Instructions for connecting tubing to tray quadrants for temperature control.

Determine the water bath temperature

Tray quadrant temperatures and pressures

Water bath and pump specifications

Coolant

Specifies approved coolant types and mixtures for the system.

Related product manuals