Operation Guide
Cool On-column Injection
19
Cool On-column Injection
For an 6890 GC with a cool on-column inlet, the 7683 Automatic Liquid Sampler
can perform injections directly onto 250-µm, 320-µm, and 530-µm columns. For
a list of consumables needed to perform these injections, see Volume 2 of your
GC Operating Manual.
When performing cool on-column injections, the injector:
Slows the carriage speed so the overall injection time increases to
500 milliseconds.
Lowers the tip of the syringe needle an additional 19 mm onto the column.
If you need to adapt your injector for cool on-column injection, change the inlet
column size, or change the septum nut or cool on-column insert, refer to your
Agilent 7683 Installation Guide and to your Agilent 6890 GC Operating Manual,
Volume 2, Inlets.