Operation Guide
Running A Priority Sample
8
Running A Priority Sample
If the Use priority parameter is on in your 6890 GC sequence, 7683 Automatic
Liquid Sampler tray gripper checks for a vial in position 100 before loading each
new vial into the injector turret. If it finds a vial there, it will load that sample
vial into the turret and run it according to the method you assigned for priority
samples. When the priority sample run is completed, it will return the priority
sample vial to tray position 99. The tray gripper will again check vial position
100, and if there is no vial there, your original sequence will continue from where
it was interrupted. For more information on how to program a priority sequence,
see your 6890 Plus GC documentation.