Operation Guide
Interrupting A Run Or Sequence
7
Restarting An Interrupted Sequence
To restart an aborted sequence from the point of interruption:
1. Resolve the problem that caused the interruption.
2. Identify the last sample vial that was run successfully. (Check the vial
number of the last successful chromatogram.)
3. Assign the first sample vial run to be the vial number after the last successful
one.
4. Start the revised sequence from your controlling device. (The
Agilent ChemStation allows you to run a partial sequence to keep the
original sample numbers.)
Following an interrupted sequence, the tray will try to return a vial it was holding
in its gripper arm to its proper tray position before beginning or resuming the
sequence. Occasionally, you may need to manually return the vial to the tray
before the sequence will start.