Operation Guide
Controlling Sample Vial Temperatures
24
Water bath and pump specifications
The water bath and pump system used to control the sample vial temperatures
must meet these specifications.
The components must meet national standards for safety requirements, be
suitable for unattended operation, be suitable for continuous operation, and
be controllable for high-temperature protection.
The minimum cooling power required for the water bath is 100 W at bath
temperatures of 5°C.
If you use a built-in pump, it must be suitable for external circulation of
liquid and for connection of 1/4-in. id (6.35 mm) tubing.
If you use a pressure pump, it must maintain a pressure between 1.5 and
2.5 psi.
If you use a suction pump, the pump vacuum cannot exceed –4 psi.
Coolant
Use distilled water as your cooling fluid. If desired, you may add a maximum of
3% ethylene glycol to the distilled water to prevent freezing.