Cleaning Instructions
Air Therapy Internal Mattress Components
1. Prepare the disinfectant according to the manufacturer’s recommendations.
2. Prepare a separate bucket of warm, fresh water to be used for rinsing the equipment after
cleaning/disinfecting procedures are completed as instructed.
3. Using a clean cloth dampened with the disinfectant solution, wipe all internal mattress surfaces,
including the air cells, and allow to remain wet for the manufacturer’s recommended contact time.
4. Rinse all surfaces of the air cells with fresh water and clean cloth to remove chemical and organic
residue.
5. After cleaning, dry the internal air cells with a clean, dry cloth.
6. After all mattress components are dry, reinstall the top cover.
Store the mattress in a “clean” environment until the next use.
Cleaning Blood and Other Excretions
Blood and other excretions should be wiped up while wet, if possible. These substances are more
difficult to remove once they have dried to surfaces. Dried blood and other excretions are to be removed
using ample disinfectant solution in order to moisten the substance and make it easier to clean.
Control Unit Cleaning Instructions
NOTE
: Hand clean only. DO NOT place in sterilization room or chamber.
1. Personal Protective Equipment should be used as directed by the Material Safety Datasheet for the
disinfectant.
2. Prepare the disinfectant according to the manufacturer’s recommendations.
3. Turn off the control unit and disconnect it from all electrical power to avoid electrical shock.
4. When cleaning the control unit avoid excessive moisture, especially in areas where there are electrical
connections and components, to prevent damage.
5. All surfaces of the control unit and the power cord are to be wiped using a coarse cloth, dampened
with the disinfectant solution prepared as directed by the manufacturer’s recommendations, to
remove organic material and visible soil.
6. Allow the control unit to remain wet with disinfectant solution for the manufacturer’s recommended
contact time.
7. Rinse all surfaces of the control unit with fresh water and use a clean cloth to remove chemical and
organic residue.
After cleaning, all surfaces are to be dried with a clean, dry cloth.
Filters
NOTE:
To keep equipment working efficiently and effectively for an extended time period, it is essential
to make sure the equipment is cleaned regularly and well maintained.
NOTE
: DO NOT attempt to clean HEPA filters, as they are to be removed and replaced only.
To clean open cell foam air filter ONLY:
1. Remove the air filter located on the back of the control unit.
2. Clean with prepared disinfectant solution and allow to air dry.
3. Reinstall the filter when dry.