Troubleshooting
ELECTRICAL SHOCK HAZARD WARNING:
To avoid electrical shock, DO NOT open
the control unit. Refer servicing to qualified personnel only.
CAUTION: Only authorized personnel should engage in the troubleshooting process.
Troubleshooting by unauthorized persons could result in personal injury or equipment
damage.
NOTE: If troubleshooting process does not solve the problem, please contact an Agiliti representative
for further assistance.
Problem Troubleshooting
Mattress not inflating
Check that the hoses are not punctured, kinked or
disconnected.
Check for proper connections from the hoses to the control
unit. Make sure they are secure.
Check air filters on back of control unit and clean if
necessary.
Ensure CPR function is in the closed position.
Power loss
Check the ON/OFF switch.
Check power cord for any damage.
Unplug the Control Unit and check fuses located near the
main ON/OFF switch. Replace fuse/s as necessary.
Ensure unit is plugged into a working power outlet.
Reduced air flow
Check for possible obstructed/contaminated foam filter.
Replace as necessary.
Check for possible obstructed/contaminated HEPA filter.
Replace as necessary.
Excessive control unit
noise
Check for possible obstructed/contaminated foam filter.
Replace as necessary.
Check for possible obstructed/contaminated HEPA filter.
Replace as necessary.
Odor
Check for excessive contamination on HEPA filter. Replace
as necessary.
Fuses
To open: insert flat head screwdriver into the center slot and turn counterclockwise.
28
IFU0000073 Immerse User Manual