EasyManua.ls Logo

agiliti Immerse - Filter Maintenance and Replacement

agiliti Immerse
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
HEPA Filter
Information
Each Unit is equipped with one open cell foam air filter and one high efficiency particulate air
(HEPA) filter.
HEPA filters aid in contamination control for facilities.
HEPA filters do not filter out gases and odor molecules such as chemical vapors and cigarette
smoke.
HEPA filter must be used in conjunction with the manufacturer supplied foam filter.
DO NOT attempt to install HEPA filter without the original equipment foam filter. Ensure factory
open cell foam filter is completely dry before placing HEPA filter in unit.
The HEPA filter is to be installed in applicable models and pumps in conjunction with the next
factory filter cleaning procedure.
DO NOT MODIFY HEPA filter for installation. Order filter from Agiliti to receive filter with correct
dimensions.
DO NOT attempt to clean HEPA filters, as they are to be removed and replaced only.
Installation
Installation and/or handling of HEPA filter requires PPE. Service personnel should wear a mask,
gloves, and protective clothing to avoid exposure to possible contaminants.
Tools are not required for installation of filter.
Ensure unit is off and/or removed from power source prior to servicing or replacement of filter
or HEPA filter.
1. Locate and remove factory filter on the top and back of the case.
2. Filter is easily removed and reinserted through the gap in the back of the case.
3. Ensure HEPA filter is clean.
4. Place HEPA filter on front of factory foam filter.
5. Re-install filters in control unit case.
Open Cell Foam Air Filter
The foam air filter on the back of the control unit must be cleaned weekly with disinfectant solution
(see cleaning section). Replacement of the foam filter is recommended every 6 months.
27
IFU0000073 Immerse User Manual