6. ELECTRONIC FUNCTIONS
Backrest Section Up & Down
Auto-Contour
ExitSafe
Ingress / Exit Height
Hi/lo Up & Down
Legrest Section Up & Down
Trendelenburg +/-
Chair Position
6.2 ATTENDANT CONTROL PANEL ACP OPERATION
WARNING
• Ensure the patient, caregivers and/or equipment is positioned
to avoid being trapped or crushed during bed adjustment
• Ensure the power lead is not stretched excessively
• Ensure the handset and electrical leads cannot become
entangled with moving parts of the bed
• DISCONNECT the power lead from the mains and stow before
transporting the bed
• If the power lead shows any sign of damage, replace immediately
• Ensure safety key is removed when handset is not in use
6.1 HANDSET OPERATION
FL250 Beds are powered by four electric IP54 rated actuators, which are
controlled by the Handset or Attendant Control Panel. Adjustment occurs only
while a button is pressed. Adjustment will stop when the button is released,
or the selection reaches the end of its range.
Auto-Contour: Conveniently positions patient into cardiac chair
position for therapeutic benefit and assists in preventing skin
shear/patient sliding down the bed when raising backrest
Backrest Section: Controls the backrest angle from horizontal
up to a maximum of 60º
Knee Break Section: Allows for knee positioning up to 35° for
increased user comfort and pressure therapy relief. The ratchet
legrest can also be enabled by hand to further increase comfort
Hi/lo with SmartDescent: Positions the bed to the desired
height from 110-740mm. The bed automatically halves its rate of
descent from 300mm to lowest position to minimize the risk of
entrapment or damage to low lying items such as mattress pumps
Trendelenburg +/- : Facilitates overall +ve or -ve tilting
of the bed from -14º to 14º from horizontal
Patient Lock Out: Locks handset out of specific function
Underbed Light: Creates a subtle glow on the floor near the bed
end for carers and users during the night