21
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare
environments and so on.
Warning: Don’t near active HF surgical equipment and the RF
shielded room of an ME system for magnetic resonance imaging,
where the intensity of EM disturbances is high.
Warning: Avoid adjacenting to or stacking with other equipments
when using this equipment because it could result in improper
operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating
normally.
Warning: Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could
result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper
operation.”
Warning: Portable RF communications equipment (including
peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the Upper Arm
Electronic Blood Pressure Monitor (M/N: FC-BP131), including cables
specified by the manufacturer. Otherwise, weakening of the
performance of this equipment could result.
If any: a list of all cables and maximum lengths of cables (if
applicable), transducers and other ACCESSORIES that are replaceable
by the RESPONSIBLE ORGANIZATION and that are likely to affect
compliance of the ME EQUIPMENT or ME SYSTEM with the requirements
of Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may
be specified either generically (e.g. shielded cable, load impedance)
or specifically (e.g. by MANUFACTURER and EQUIPMENT OR TYPE
REFERENCE).
If any: the performance of the ME EQUIPMENT or ME SYSTEM that was
determined to be ESSENTIAL PERFORMANCE and a description of
what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is
lost or weakened due to EM DISTURBANCES (the defined term
“ESSENTIAL PERFORMANCE” need not be used).