3.6.1 All necessary instructions for maintaining BASIC SAFETY and
ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances
for the excepted service life.
3.6.2 Guidance and manufacturer’s declaration -electromagnetic
emissions and Immunity
Table 1
Table 2
Guidance and manufacturer’s declaration - electromagnetic emissions
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Compliance
Guidance and manufacturer’s declaration - electromagnetic Immunity
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact±2 kV,
±4 kV, ±8 kV,
±15 kV air
±8 kV contact±2 kV,
±4 kV, ±8 kV, ±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
Power supply lines:
±2 kV 100 kHz
repetition frequency
Power supply lines:±2 kV
100 kHz repetition
frequency
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
0% 0.5 cycle At 0º,
45 º, 90 º, 135 º, 180 º,
225 º, 270 º and 315 º
0% 1 cycle And 70%
25/30 cycles Single
phase: at 0 0%
300 cycle
line(s) to line(s):
±1 kV.
line(s) to line(s):±1 kV.
0% 0.5 cycle At 0º, 45 º,
90 º, 135 º, 180 º, 225 º, 270 º
and 315 º 0% 1 cycle And
70% 25/30 cycles Single
phase: at 0 0% 300 cycle
IEC 60601-1-2
Test level