Page 72
13 Specification
PSA3605 limits, where different, are shown in square brackets [ ] after the PSA6005 limits.
13.1.1 Frequency Measurement
Frequency Span
10 MHz to 6000 MHz [10 MHz to 3600 MHz] in one range for Normal
sweep mode
30 MHz to 6000 MHz [30 MHz to 3600 MHz] in one range for DC (data
Centre frequency plus Span, or Start frequency plus Stop frequency
5990 MHz [3590 MHz] for Normal sweep mode
27 kHz, or Zero Span with demodulation
Reference Frequency Accuracy for Start, Stop and Centre frequencies.
Reference Frequency Accuracy
Better than ± 1 ppm at 20
o
C
Better than ± 1 ppm over 10
o
C to 30
o
C
Better than ± 1 ppm per year
Phase Noise (RBW = 3 kHz, VBW = 1 kHz and Positive Detector)
Carrier Frequency = 1 GHz
-83 dBc/Hz at 30 kHz offset
-99 dBc/Hz at 100 kHz offset
-116 dBc/Hz at 1 MHz offset
Carrier Frequency = 3 GHz
-94 dBc/Hz at 30 kHz offset
-109 dBc/Hz at 100 kHz offset
-117 dBc/Hz at 1 MHz offset
Carrier Frequency = 6 GHz
-83 dBc/Hz at 30 kHz offset
-97 dBc/Hz at 100 kHz offset
-114 dBc/Hz at 1 MHz offset
Resolution Bandwidth and Video Bandwidth
Selectable as 300Hz to 10MHz in 1:3:10 sequence, or Auto
Selectable as 300Hz to 10MHz in 1:3:10 sequence, or RBW Tracking
Frequency Markers
0.1 kHz at all frequencies
1/270
th
of Frequency Span ± 10Hz + Reference Frequency Accuracy
Display of absolute and difference frequencies for both markers
Frequency Counter
Selectable frequency counter at the position of the active marker
Selectable as 1kHz, 100Hz or 10Hz.
Reference frequency accuracy +/- 1 count (at > noise level + 25 dB)
Channel Markers (only with Option U02)
Channel markers are vertical lines on the screen based upon frequency
values within a file. They are a special case of Limit Patterns.
Up to 49 markers per file. Two files can be displayed in differing colors.