1. General Information – Read this first
1. Type of equipment and intended use:
Pluggo is used for safe removal of original caps from vacuum collection tubes. It is
intended for use in laboratory environment under normal in-house conditions.
Clinicon AB cannot accept any responsibility for any loss or damage, however caused,
arising from a faulty or incorrect use of the instrument.
2. Tube Handling
Pluggo is capable of handling 5ml, 7ml, 7ml long and 10ml Vacuum tubes of the same
brand.
Note: Each tube size has to be loaded in its corresponding carousel size with different
colour marks.
Note: Never load de-capped tubes in carousel. Pluggo will damage tubes without caps!
Note: Never change brand without consulting Pluggo technical support.
3. Carousels
Different tube sizes are handled by use of different carousels with corresponding colour
marks.
Carousel colour Tube size Volume
Note: All tubes placed in a carousel must be of the same type and size. Otherwise,
damaged tubes may result.
Each carousel is capable of loading up to 24 tubes.