4. Cleaning Instructions
Due to the handling of tubes with blood samples, it is important to clean all potential
infectious parts on a daily basis. Use clinic/laboratory approved disinfectant solution.
Never use Bleach!
Note: Always use protective gloves when cleaning equipment.
1. Remove chute from the unit for cleaning procedure using disinfectant solution or in
auto-clave equipment.
2. To remove the de-capper head:
Remove any carousel
Press Start to let de-capper head move outwards.
Press Stop button. This will let de-capper arm to stop in out position
Push de-capper head and turn left (90 degree counter clockwise). Se figure 5.
Pull out the head gently from arm.
The de-capper head is cleaned with disinfectant solution or in auto-clave.
Warning: The sharp needle will be exposed during removal. It presents a potential hazard
from possible contact with infected blood samples. Take all necessary precaution when
handling de-capper head or chute.
De-capper head To remove head Push and turn left 90 degree
Figure 5: Removal of de-capper head
3. Clean unit and cover by using disinfectant solution on a cleaning cloth.
Note: Never use spray solution on Pluggo unit! It may affect the mechanical function.