DS100 Mini oscilloscope
User Manual www.alientek.com
3
DS100 Mini Oscilloscope User Manual
Notices
1 This document applies to firmware version V3.7 and hardware version V2.4.
Other versions will be slightly different.
2 Only when the hardware charging indicator light goes off or the charging icon
disappears will it be fully charged.
3 Due to the difference in the size of the MCX pin of the alligator clip and the
oscilloscope probe (the inner pin of the alligator clip is larger), please do not
attach the alligator clip to the CHA/CHB for use. The alligator clip is only used at
the OUT channel ( waveform output ), so as to avoid the use of the alligator clip
with the large MCX hole of the oscilloscope analog channel, resulting in poor
contact of the oscilloscope probe.
4 Please refer to the section on vertical systems to enable channel B.
5 The four keys of the oscilloscope are fragile. In order to avoid damage (collapse),
please press them as gently as possible and avoid pressing them vigorously.
6 When the contact between the oscilloscope probe and the oscilloscope is
loose/bad, you can use fingernail/forceps to push the 6 reeds of the outer ring of
the probe MCX seat slightly out (do not dial too much, otherwise it can not be
inserted) to increase the contact pressure and restore the contact well.