1. Plug dual connector end of wires into
Channel 1 or Channel 2 jack and pin
plugs into Smart Probes. Figure 1.
2. Apply Probe Electrode Pads (PEPS™) to
Probes by placing clean Probe tip into
PEP™ in PEP™ case. PEP™ case opens
with label on bottom. Figure 2.
3. Press Power on.
4. Set Frequency to 0.5 Hz or desired setting. For joint problems (e.g.,
shoulders, elbows, wrists, fingers, hips, knees, ankles, toes) use 100 Hz for
10 - 20 seconds immediately followed by 0.5 Hz. Try 1.5 Hz if 0.5 Hz is not
effective.
5. Set Timer to Probe setting.
6. Set Current on Channel in use to 6 (600 µA). Decrease immediately if
uncomfortable. When used on or near the head, immediately decrease
the current if dizziness or nausea develops. This may happen initially or
several minutes into treatment. Decreasing the current will immediately
relieve these unpleasant feelings.