Available currents, use and starting settings -----------------------------------------------------------------------------
PURE (monopolar pure cut without coagulating effect).
It is suitable to obtain each pure cut: uterine conization in gynaecology, papillotomy in flexible endoscopy, cut under liquid
for minor isteroscopy and so on. It gives better results by using thin electrodes:
Needle, knife electrodes (Starting settings from 3-4W).
Thin loop electrodes or very thin needle electrodes: SAD, SAD/1, SAD/2, SAD/3 from 0.1 to 0.4 mm. (Starting setting from 3-4 W).
LLETZ/LEEP electrodes for conization (Starting setting from 10 W).
Electrodes for papillotomy (Starting setting from 20/30 W).
P
PULSED
(monopolar pulsed pure cut- fast).
It is identical to the PURE cut and is usable with same electrodes and powers, but it is more suitable to obtain very fine cuts.
BLEND (monopolar coagulating cut).
It is suitable to obtain a properly coagulating cut and it is usable with the identical electrodes and settings of the PURE cut.
It is also suitable to perform polipectomies (starting setting from 20/30W).
B
PULSED
1 (pulsed monopolar coagulating cut - fast)
It is identical to the BLEND cut and is usable with same electrodes and powers, but it is more delicate.
B
PULSED
2 (pulsed monopolar coagulating cut - slow)
It is identical to the BLEND cut, but it is especially suitable to perform polypectomies (starting setting from 20/30W).
MICRO (delicate coagulation with minimum sparking).
It is suitable to perform all micro coagulations (i.e. the epilation) and the coagulation of not bleeding tissues by using
following electrodes:
AID: insulated thin needle for epilation (starting setting from 1-2 W).
SAD, SAD/1, SAD/2, SAD/3: very thin not insulated needles from 0.1 to 0.4 mm. (starting setting from 3-4 W).
Ball electrodes 2-2,5mm (starting setting from 15-20 W).
Ball electrodes 3-4mm (starting setting from 20-30 W).
M
PULSED
(Temporized delicate coagulation with delivery time adjustable from one tenth of second (1) to 1 second (10).
It is identical to the MICRO coagulation, but grants to its very short delivery time is the best to perform all
microcoagulations.
FULG (macro coagulation “fulguration” with high sparking)
It is suitable to coagulate all both bleeding and not bleeding tissues by using following electrodes:
Needle, knife or loop electrodes, ball electrodes and surgical forceps (starting setting from 20-30 W).
LLETZ/LEEP electrodes for conization (starting setting from 20-30 W).
Loops for flexible endoscopy (starting setting from 20-30 W).
It is perfect to perform a cut with strong coagulating effect by using needle or knife electrodes (starting setting 20-30 W).
FULG
PULSED
(Pulsed macro coagulation “fulguration)
It is identical to the FULG coagulation, but it is more suitable for same uses if users like a more delicate result.
BIPOLAR (bipolar coagulation).
It is suitable to perform bipolar coagulations with following instruments:
Bipolar forceps (If provided with tips 0,5/1mm, starting setting 1-2W) (If provided with tips 1,5/2mm, starting setting 4-5 W)
Needles for turbinals or similar uses (starting setting from 20 – 30 W).
Forceps for laparoscopy (starting setting from 20 – 40 W)).
Adaptors and connecting cables for not standard accessories - Use of the RD/5 adaptor -------
MONOPOLAR CABLES AND ADAPTORS:
ALSA cables (Specify model and connector of the instrument).
ALSA cables for electrodes for flexible endoscopy:
CEP/3: cable (3 mt) for electrodes with male plug Ø 3mm.
CEP/3-5: ditto (5 mt).
CEP/4: cable (3 mt) for electrodes with female connector Ø 4mm.
CEP/4-5: ditto (5 mt).
RD/5: Universal adaptor to use all cables.
BIPOLAR CABLES AND ADAPTORS:
ALSA cables (Specify model and connector of the instrument).
Adaptors to use all following cables:
RD/BF: Cables with Valleylab connector or with 2 flying plugsØ 4 mm.
RD/BF1: Cables with coaxial connector Ø 12,5 mm – ERBE, STORZ).
RD/BF2: Cables with coaxial connector Ø 9 mm – Martin).
USE OF THE RD/5 ADAPTOR
1.To use cables with plug Ø 4 mm.
Unscrew the knob, take away the first metallic part and insert the cable.
2.To use cables with plug Ø from 2 to 3 mm.
Unscrew the knob, take away the first metallic part, insert the plug and screw the knob again.
3.To use cable with plug Ø from 4,5 to 8 mm. (ERBE, STORZ, VALLEYLAB, CONMED, others)
Unscrew the knob, take away both metallic parts , insert the plug and screw the knob again