Sequoia / Tahoe Series User Manual California Instruments
M447352-01 REV-A 101
8. Navigate to the CURRENT menu in the OUTPUT PROGRAM screen and adjust the output current
value.
9. Navigate to the FREQUENCY menu in the OUTPUT PROGRAM screen and adjust the output
frequency value.
10. The output could be turned on with the front panel OUTPUT switch.
Grid Simulator Mode: (Applicable only for Sequoia series)
1. Navigate to the PHASE NUMBER menu in the CONFIGURATION screen and select the output
phase number: either One-Phase or Three-Phase.
2. Navigate to the MODE menu in the CONFIGURATION screen, and select the output voltage
mode: either AC, DC, or AC+DC.
3. Navigate to the RANGE menu in the CONFIGURATION screen and select the output range: either
Low-Range or High-Range.
4. Navigate to the REGULATION menu in the CONFIGURATION screen and select the output
voltage/current regulation: either Constant Voltage/Constant Current or Constant Voltage/Current
Limit.
5. Navigate to the VOLTAGE menu in the OUTPUT PROGRAM screen and adjust the output voltage
value.
6. If the AC+DC voltage mode had been selected, navigate to the DC OFFSET menu in the
CONFIGURATION screen, and adjust the DC component of the output voltage.
7. Navigate to the SOURCE CURRENT menu in the OUTPUT PROGRAM screen and adjust the
output Source Current value.
8. Navigate to the REGENERATIVE CURRENT menu in the OUTPUT PROGRAM screen and adjust
the Regenerative current value.
9. Navigate to the Regenerative Control Settings menu in the CONFIGURATION screen and adjust
the Regenerative Under Volt limit and Over Volt limit value.
10. Navigate to the FREQUENCY menu in the OUTPUT PROGRAM screen and adjust the output
frequency value.
11. The output could be turned on with the front panel OUTPUT switch.
eLoad Mode: Current Programming (Applicable only for Sequoia series)
1. Navigate to the PHASE NUMBER menu in the CONFIGURATION screen and select the output
phase number: either One-Phase or Three-Phase.
2. Navigate to the MODE menu in the CONFIGURATION screen and select the output voltage mode:
either AC or DC.
3. Navigate to the RANGE menu in the CONFIGURATION screen and select the output range: either
Low-Range or High-Range.
4. Navigate to the SOURCE CURRENT menu in the OUTPUT PROGRAM screen and adjust the
source current value.
5. Navigate to the PHASE SHIFT menu in the OUTPUT PROGRAM screen and adjust the Phase
Shift value.
6. Navigate to the SYNC Settings menu in the CONFIGURATION screen and adjust the Sync
Voltage, Sync Phase, and Sync Frequency values.