Sequoia / Tahoe Series User Manual California Instruments
154 M447352-01 REV-A
2. Scrolling to the OUTPUT PROGRAM screen with the encoder and depressing the encoder switch.
The UP arrow button will return to the previously selected screen menu (in this case the HOME Screen).
The HOME button will return to the HOME screen that has the top-level menu for the sub-menu being
displayed; for the OUTPUT PROGRAM screen top-level menu, is the HOME Screen.
Figure 5-85. OUTPUT PROGRAM Screen Top-Level Menu
The following choices are available in the OUTPUT PROGRAM screen top-level menu. Functions that
accept a numeric value require that the value is within the allowed range, otherwise, an error will be
generated, and the value will not be accepted.:
Entry Description
Settings
VOLTAGE Programs the output voltage in RMS value, V(RMS), when in
AC-mode and DC-mode, and the AC component when in
(AC+DC)-mode. In (AC+DC)-mode, the DC component is set
separately using the DC OFFSET selection field (below), or
through the Dashboard screen. In DC mode, negative values can
also be entered. The default is zero.
Figure 5-86. Voltage Settings
FREQUENCY Programs the output frequency in Hz when in AC mode. If the unit
is in DC mode, the value for FREQ will be set to DC and cannot
be changed until AC mode is selected. When in AC-mode, the
frequency can be changed from 16 Hz to 905 Hz (depending on
the options). The default is 60 Hz.