Figure 5-210. PROFILES Menu, NAME Sub-Menu
USER CURRENT LIMITS Sets soft limits for the minimum and maximum output voltage to
which the unit could be programmed using the front panel or
remote digital interface; the default is full-scale.
Figure 5-211. User Current Limit Settings
DEFAULT SCREEN Selects whether the Default screen (showing only voltage and
current amplitude) is enabled and configures its operational
characteristics; the defaults are Default screen enabled, 10-
second timeout.
Timeout Interval: Select the time, in seconds, for how long a
screen must be inactive before the Default screen is displayed.
Figure 5-212. Default Screen Settings
PONS The PONS menus allow setting the conditions that would be
present after power up; The AC input has to be cycled off/on for
a change in a PONS setting to take effect. The functions and
parameters have the same programmability as described in the
menus of the OUTPUT PROGRAM screen.
CAUTION!
The PONS menus allow selecting that the output would be turned on and programmed
to a high voltage when the unit is initially powered up. Ensure that suitable protection is
provided to prevent accidentally energizing the load. The factory-default setting is with
the output off and programmed to zero to provide the safest start-up condition.