Sequoia / Tahoe Series User Manual California Instruments
M447352-01 REV-A 237
Figure 5-255. CONTROL INTERFACE REMOTE INHIBIT
Menu
Latching: A TTL logic signal at the external Remote Inhibit input
latches the output in the shutdown state; when the output is turned
off, it is programmed to zero volts, and the output relays are opened;
this state could only be cleared by the remote digital interface
SCPI command, OUTPut:PROTection:CLEar;
Live: The output state follows the state of the external Remote
Inhibit input, turning the output on/off.
Low/High: Selects the logic level of the Remote Inhibit signal that
would cause the output to be turned off: either a logic low or
contact closure, or a logic-high or open circuit.
Off: The power source ignores the external Remote Inhibit input.
AMBIENT STATUS Configures the Ambient led status feature to turn on/off.
Figure 5-256. Ambient Status LED setting Screen
5.13 System Settings Screen
The SYSTEM SETTINGS screen provides information on versions of firmware and which options are
installed. It also allows for selecting the language used for the display, performing calibration of the touch-
screen, and viewing hardware limits settings.
The top-level menu of the SYSTEM SETTINGS menu is shown in Figure 5-257. It can be reached in one
of two ways: