Sequoia / Tahoe Series User Manual California Instruments
266 M447352-01 REV-A
Figure 8-21: List Menu with Empty Buffer
4. Tap the ADD field to enter the ADD sub-menu; refer to Figure 8-22.
Figure 8-22: Current Transients Type Selection for
eLoad – Current Programming Mode
5. Tap the CURRENT SURGE/SAG selection button, refer to Figure 8-23. Enter the following
parameter values:
T(ime): 0.083 sec; the value is entered as seconds, with a minimum time resolution of 0.001 sec;
A(mpere): 80 C; the surge current value;
To A(mpere): 0 C; the current value following the surge;
Repeat: 0; the number of times to repeat this transient event (not the entire transient list, as
describe in Step 10, below);
Function: Sine; output waveform
Trig: blank (no selection); not used in this example
Delay: 0.067 sec; time interval to remain at To A(mpere) level.