7
T-900 Series Pump Operation Manual
Device-Under-Test Pressure Port Connections
In order to adapt the different connection threads of the device-under-test, the pressure hose can be fitted with different adapters. Please use a suitable sealing
gasket or Teflon tape as applicable for the thread type.
! CAUTION: Do not use teflon tape with BSP or CPF threads; this may damage your hand pump.
! CAUTION: The T-900 Series hand pump must not be soiled, or come into contact with fluids or aggressive media.
! CAUTION: To prevent leaks, tighten the tube fitting or CPF connection to a maximum torque of 15 N-m = 11 lb-ft.
Actuate the Shuttle Valve (Combination Models Only)
Verify that the shuttle valve is positioned to provide pressure or vacuum. Use a pen or a small screwdriver for this purpose.
The encasement of the switch is intended to help prevent unintentional actuation. (Only applies to models T-965, T-975, and T-975-CPF.)
Operating and Safety Instructions 6