84 / 101 NTA47230-300A6
TC-TM 66XX
Single ramp signal configuration?
The figure below shows the type of single ramp that can be generated along with its parameters:
The LOW level and HIGH level levels are expressed:
As a percentage of the range if the scaling mode is ON.
In Volt or in temperature units if the scaling mode is OFF and according to the type of value emitted (voltage or
temperature emission).
The Total Duration corresponds to the incrementation time it takes to go from Low Level to High Level (and vice versa
with decrementation). It is given in seconds and is limited to a maximum of 1000s.
The Time frame corresponds to the amount of time you can have between pressing the emission start key and the actual
starting of generation. It is expressed in seconds and the max time is limited to 1000s.
Use the F2 key to access the CONFIGURATION/RAMP menu. Attention: the appropriate function mode must be
programmed (INCREASES mode) to access the CONFIGURATION/RAMP/SINGLE menu.
In the CONFIGURATION/RAMP/SINGLE menu, use the F2 key to go to the next field.
Use the navigation keys to enter the value:
As a percentage of the gauge if the scaling mode is ON.
and to increase or decrease the value
And to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press VAL to save the parameters.
To quit the menu without saving, press CLEAR.
Programming parameters in Single Ramp mode
T0
Tdt
T0
Td
Tdt
T0
Tdt
NH
NB
Td
To : time pressing the generation start key
Tdt : Total Duration
Tdt : Total Time
NB : Low Level
NH : High Level