TECHNICAL SPECIFICATIONS
94 / 101 NTA47230-300A6
TC-TM 66XX
Sensor
Scope of measurement
Resolution
Precision / 1 year
K
- 250 to - 200°C
- 200 to - 120°C
- 120 to - 50°C
-50 to + 1 372°C
0.20°C
0.10°C
0.05°C
0.05°C
0.90°C
0.3°C
0.02 % R + 0.12°C
0.02 % R +0.11°C
T
- 250 to - 200°C
- 200 to - 50°C
- 50 to + 400°C
0.2°C
0.05°C
0.05°C
0.80°C
0.25°C
0.02 % R +0.09°C
J
- 210 to - 200°C
- 200 to - 120°C
- 120 to + 60°C
+ 60 to + 1 200°C
0.05°C
0.05°C
0.05°C
0.05°C
0.30°C
0.25°C
0.020 % R + 0.11°C
0.020 % R + 0.09°C
E
- 250 to - 200°C
- 200 to - 100°C
- 100 to + 450°C
+ 450 to + 1 000°C
0.1°C
0.05°C
0.05°C
0.05°C
0.55°C
0.20C
0.020 % R + 0.07°C 0.020 %
R + 0.05°C
R
- 50 to + 150°C
+ 150 to + 550°C
+ 550 to + 1 768°C
0.50°C
0.20°C
0.10°C
0.95°C
0.40°C
0.020 % R + 0.30°C
S
- 50 to + 150°C
+ 150 to + 550°C
+ 550 to + 1 768°C
0.5°C
0.2°C
0.1°C
0.85°C
0.020 % R + 0.4°C
0.020 % R + 0.3°C
B
+ 400 to + 900°C
+ 900 to + 1 820°C
0.2°C
0.1°C
0.95°C
0.50°C
U
- 200 to - 100°C
- 100 to + 600°C
0.05°C
0.05°C
0.35°C
0.20°C
L
- 200 to - 100°C
- 100 to + 900°C
0.05°C
0.05°C
0.30°C
0.20°C
C
- 20 to + 900°C
+ 900 to + 2 310°C
0.1°C
0.1°C
0.30°C
0.020 % R+ 0.15°C
N
- 240 to - 190°C
- 190 to - 110°C
- 110 to 0°C
+ 0 to + 1 300°C
0.2°C
0.1°C
0.05°C
0.05°C
0.60°C
0.25°C
0.15°C
0.020 % R + 0.07°C
Platinum
- 100 to + 1 400°C
0.05°C
0.3°C
Mo
0 to + 1 375°C
0.05°C
0.020 %L + 0.10°C
NiMo/NiCo
- 50 to + 1 410°C
0.05°C
0.020 %L + 0.35°C
D
+ 0 to + 310°C
+ 310 to + 1000°C
+ 1000 to + 1800°C
+ 1800 to + 2000°C
+ 2000 to + 2315°C
0,1°C
0,05°C
0,05°C
0,05°C
0,05°C
0,40°C
0,40°C
0,025 % R +0,15°C
0,030 % R +0,10°C
0,05 % R
G
+ 0 to + 100°C
+ 100 to + 300°C
+ 300 to + 1100°C
+ 1100 to + 1800°C
+ 1800 to + 2315°C
0,5°C
0,2°C
0,05°C
0,05°C
0,05°C
2.5°C
0,6°C
0,3°C
0,5°C
0,85°C
The precision is guaranteed for a reference junction (JR) at 0°C.
With the use of the internal JR (except for thermocouple B), add an additional uncertainty of 0.3°C to 0°C. For the other
temperatures, it is a good idea to take into consideration the sensitivity of the thermocouple at the temperature (T)
considered, namely an additional uncertainty of 0.3°C*S(0°C)/S(T).
Temperature coefficient: < 10 % of precision/°C from 0°C to 18°C and from 28°C to 50 °C.
You can, except for thermocouple B, choose the location of the reference junction by programming it on the
keyboard:
o external at 0°C,
o Internal (temperature compensation at the device terminals).
o By programming the temperature.

Table of Contents