Section 5 – Specifications
5.1 Specifications
Range of Measurement : 0 ~30mg/㎗(Total bilirubin)
: 0 ~ 513.09 μmol/ℓ
Measurement accuracy : ±5%
Correcting Hemolysis : 0 ~250mg/㎗ Hbcv
Measuring method : Dual wavelength differential system (461nm,551nm)
Sample volume : 50 ~ 60 ㎕
Sample container : Hematocrit capillary tube (Heparinized)
Light source : 5V, 0.75Atungsten lump
Detector : Silicon photocell
Display : LCD Display 16 rows × 2 lines
Power requirements : 100~240VAC 50/60Hz, 10VA
Dimensions : 280mm(W)×230mm(D)×115mm(H)
Weight : Net 3.4kg (Main body)
Operating temperature range : +10°C ~ +40°C
Storage temperature range : 0°C ~ +55℃
Humidity : 30% ~ 75%
Dew condensation must be avoided.
Pollution Degree :
2
( IEC 1010-1, IEC 664 )
Overvoltage Category : Ⅱ ( IEC 1010-1, IEC 664)
Fuse ( In Power Unit ) : T 3.15A (IEC 127)