14
3. Temperature Coefficient : 0.2 x (Spec.Acc'y) /℃ < 21 ℃ or > 26℃.
4. Operating Temperature : 0℃ to 30℃ (≦ 80%RH) , 30℃ to 40℃ (≦ 75%RH)
AC Conversion Type : 36Ⅱ:
AC conversion are average sensing rms indication calibrated to the rms value of a sine
wave input.
36RⅡ:
AC conversions are ac-coupled, true rms responding, calibrated to the rms value of a
sine wave input . Accuracies are given for sine wave at full scale and non-sine wave
below half scale. For non-sine wave add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.