506/506B
EN
Relative Δ
Relative %
When measuring, you can use the relative (Δ) mode to
subtract the o󰀨set.
To use relative (Δ) mode, move the blink cursor of menu to Δ
position, and press the ENTER button to enter mode. In this
mode, meter records the presently reading as reference and
shows it on the secondary display. The relative (Δ) mode
subtract reference from each reading, and shows result on
the main display.
In relative (Δ) mode, the minimum range is relative (Δ)
range. For example: The ranges of measuring resistance
are 400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, and 40MΩ. If you
use the relative (Δ) mode in 4kΩ range, then the minimum
range is the 4kΩ range. You can use range is 4kΩ to 40MΩ,
cannot change to 400Ω.
To exit relative (Δ) mode, press the CANCEL button.
When measuring, you can use the relative (%) mode to
calculate the relative percent value. The relative percent
value is dene as below:
To use relative (%) mode, move the blink cursor of menu to
% position, and press the ENTER button to enter mode. In
this mode, meter records the presently reading as reference
and shows it on the secondary display. The relative (%)
mode calculate the relative percent value from each reading,
and shows result on the main display.
To exit relative (%) mode, press the CANCEL button.