(7) DCR
AccuracyRange
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
20.000MΩ 1kΩ 0.5% + 5
[2]
200.00MΩ 10kΩ 1.0% + 5
[1][2]
[1] < 50dgt rolling.
[2] When equip the external power, the accuracy
is unspecified.
Resolution
Input Protection : 30V
DC
or 30V
AC
rms
Minimum Resolution : 0.01Ω
Measuring Time : 2 seconds
D & Q
Definition : Q = 1/D = tanθ
Range : 2.000 to 2000
Minimum Resolution : 0.001
Accuracy : Basic Accuracy x (1+D)
Note : When equip the external power, the accuracy is
unspecified.
ESR
The specification of ESR is same as Resistance.
θ
Definition : θ = tan
-1
Q
Range : -90.0° to 90.0°
Minimum Resolution : 0.1°
Accuracy : 0.2% + 5
Note:
When equip the external power, the accuracy is unspecified.