12
EN
5. Replace the back cover and re-install
the 4 screws (do not over tighten) and
rubber feet.
The internal date and clock must be re-set after
replacing the battery. Please see the ‘One-Time
Setup’ section at the beginning of this manual.
TROUBLESHOOTING
Problem:
The analyzer doesn’t turn on when connected
to a battery
Cause:
n
The analyzer clamps are connected to the
incorrect poles of the battery.
n
Battery voltage is below 6V. If battery
voltage is below 6V, the analyzer will not
turn on. Charge battery and retest.
Problem:
The analyzer does not provide a similar result
when a test is repeated.
Cause:
n
The terminals are not clean or the clamps
are not making secure contact with the
battery terminals. Reconnect clamps.
Problem:
The analyzer indicates a failed battery after the
capacity test, but a passed battery after the
cranking test.
Cause:
n
A wrong value was entered as the
reference value for the capacity test.
n
The battery capacity has dropped below the
manufacturers specification, but still has
enough power to start the particular engine.
n
The battery is larger than necessary for the
starting load.
Problem:
The analyzer indicates a passed battery after
the capacity test, but a failed battery after the
cranking test.
Cause:
n
A wrong value was entered as the
reference value for the capacity test.
n
The battery is too small for the starting
application.
If you have other troubleshooting questions,
please visit the support section of our web site.