What to do if Arjo Indoor Furnishing shows hardware failure and the repeater light is on?
- PPhillip WilliamsSep 12, 2025
If the hardware failure occurs on initial power up, it indicates an internal hardware fault, and the pump should be replaced. If it occurs during normal operation after a successful start-up, select Standby, remove the power source, re-power, and if the pump fails the BIST (Built-In Self-Test), replace the pump.