What to do if Arjohuntleigh Lifting Systems Minstrel has unexpected movement?
- AamandaduranAug 9, 2025
If the Arjohuntleigh Minstrel Lifting System experiences unexpected movement, immediately release both control buttons. If releasing the buttons doesn't stop the movement, push the red Emergency Stop button and remove the battery. Use the System Failure Lowering Override to safely seat the patient before removing the hoist and taking it out of service. A qualified technician must inspect and approve the Minstrel before it is used again. This could be due to pressing two control buttons at the same time or a faulty hand control, push buttons, or electronics.