EasyManuals Logo

Arkray SPOTCHEM EZ User Manual

Arkray SPOTCHEM EZ
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Testing Procedure:
1. Add whole blood to the ll line on the centrifuge cup or serum / plasma to ll line
on serum cuvette.
2. Press the [1] key on the Main Menu to initiate a measurement.
• The table cover will open and the STAND-BY screen is displayed.
3. Press the [6] key to bring up and scroll through the species menu.
4. Press [ENTER] to select the desired species.
• Unit will go back to the STANDBY screen
5. Press the [1] key for ID.
6. Enter the test number (if desired) otherwise press [ENTER].
7. ID option will appear - enter patient number or name, press [ENTER].
• The [+/.] key is the space key and can be used to move the cursor to the next number/
letter key.
8. When using serum/plasma, place the serum cup in the serum holder.
• Skip to step 12.
9. Remove centrifuge cover by pulling backwards and upwards.
10. Place the centrifuge cup with whole blood in position on the indented area of the silver
tabs.
11. Install centrifuge cover.
12. Place the pipette tip in the Tip Hole.
13. Place the reagent strips on the reagent table.
• Reagent pads on single strips should be facing up and aligned towards the rear of the
instrument.
• Reagent pads on multi strip should be facing up and aligned towards left side of the
instrument.
14. Press [START] key to begin measurement.
• When the measurement is completed, remove used strips/centrifuge cup/cuvette and
discard them.
• Instrument is ready to run the next sample.
How to run a sample:
E02
(T): Trouble with the table Cover.
(M): Trouble with the maintenance Cover.
The measurement table was open during measurement or
warm up
The maintenance cover was opened during measurement
or warm-up
The table cover is not properly closed (foreign material may
be preventing closing)
The measurement was started without closing the
maintenance cover
Close the table cover correctly (remove obstructions)
Set the maintenance cover in the correct position
E05
(1~9): Channels with abnormalities
A check shows the optical system status has changed since
the last calibration
The white and black plate or Optical window (window plate)
is dirty
Clean the white and black plate or Optical window (window
plate) - see sections 4-2-1 and 4-3-2 in the owner’s manual
E11
(7~9): Single Type (M): Multi type
Channel with no Reagent strip in either case. Display is made only in
calibration. Usually nothing is displayed during measurement.
Reagent strip is not set
The bar code was not correctly read due to displacement or
bending of the Reagent strip
The Reagent strip is not set in the channel necessary for
calibration
Set Reagent strips correctly
When calibrating, swipe all necessary numbers on card
E13
(7~9): Single Type (M): Multi type
Channel with a different Reagent strip in either case.
The bar code was not correctly read due to displacement or
bending
The wrong reagent strips was set for calibration
Calibration was started for an item before it was completed
for another item
Set the Reagent strip correctly
Return to the main menu because changing the item
mid-calibration is not allowed when doing the calibration
E14
(7~9): Single Type (M): Multi type
Channel has used Reagent strip in either case.
Reagent strip is already used
The Reagent strip pad is colored because of an old Reagent
strip of inappropriate preservation
The Reagent pad is dirty
The bar code was not correctly read due to displacement of
bending of the Reagent strip
Use a new Reagent strip
Set the Reagent strip correctly
E17
A sample vessel is misplaced
There is no sample or an insufcient amount of sample
A centrifuge tube was used in calibration by calibrator kit
Re-set the sample correctly
Conrm the amount of the sample
Be sure to use cuvettes in calibration by calibration kit
E32
Reagent strip has expired.
Clock is not correctly adjusted
Clock is not correct because the power was not connected
for a long time
When using a Reagent strip from new lot, the corresponding
magnetic card to the lot was not inserted
Use a new Reagent strip
Adjust the clock
Insert a new magnetic card
E38
The tip waste receptacle is not correctly set Set the tip waste receptacle correctly
E40
(1~9): Channels with abnormalities
Failed to draw the sample and drop it to the Reagent
correctly due to lack of sample amount or absorbing brin
Sample adhered around the dropping hole
Check the sample amount and remove brin if necessary
Wipe off the sample adhered around the dropping hole
Measure with a different tip
T25
(1-5, 7-9): Channels with abnormalities
White plate or black plate is dirty Clean the white plate or black plate. (Note: the case in
which trouble is detected when the power is turned on, it
is necessary to turn on the power again after maintenance
since measurement cannot be started)
T03
(1-5, 7-9): Channels with abnormalities
The pressure does not increase since the rubber plate is
dirty or deformed
The tube or pipe is clogged
The tube connecting the nozzle with the pump is
disconnected
The tube connecting the pump with the pressure detecting
sensor is disconnected
Clean the rubber plate
Check the tube
If the same trouble occurs, contact your distributor (Note:
in cases where the abnormality is detected when the power
switch is turned on, it is necessary to turn on the power
again after maintenance because the measurement could
not be started)
E02 Cover Open
OK (ENTER)
E05 Optical Error
(9 6 2) OK (ENTER)
E11 No Strips
( M) OK (ENTER)
E13 Wrong Strips
( 87 ) OK (ENTER)
E14 Used Strips
(9 ) OK (ENTER)
E17 No Sample
(CNTRFG) OK (ENTER)
E32 Invalid Strips
( 7 M) OK (ENTER)
E38 Tip Case Not Set
OK (ENTER)
E40 Sample Drop
( 7 4 1) OK (ENTER)
T25 Barcode Sensor
(8 5 3) OK (ENTER)
T03 Tube Pressure
OK (ENTER)
Troubleshooting
Description and Error Message Conditions and Causes Corrective Action
Centrifuge Cover
Place the Pipette Tip in Holder
Place Test Strip on Reagent Table
ll line
max 500 �L
ll line 250 �L
ll line
min 100 �L
Contact scil animal care toll free at 877.724.5838
Contact scil animal care toll free at 877.724.5838
SPOTCHEM
EZ Quick Instruction Guide SPOTCHEM
EZ Quick Instruction Guide
scil animal care company • 151 N. Greenleaf St. Gurnee, IL 60031 • P: 877-724-5838 • www.scilvet.com scil animal care company • 151 N. Greenleaf St. Gurnee, IL 60031 • P: 877-724-5838 • www.scilvet.com
Whole Blood – uses centrifuge cup
1.Using a lithium heparinized collection tube (recommended method)
• Aspirate sample into syringe (no anticoagulant)
• Remove needle from syringe
• Dispense sample into lithium heparin collection tube and mix sample
• Transfer sample into centrifuge cup
2. Direct transfer to centrifuge cup
• Aspirate sample into syringe (no anticoagulant)
• Remove needle from syringe
• Dispense sample directly into centrifuge cup
• Re-install orange cap on centrifuge cup
• Mix sample in centrifuge cup; verify sample is completely heparinized
before proceeding (minimum of 2 minutes)
Plasma/Serum – uses cuvette
• Prepare plasma/serum sample
• Transfer sample into cuvette tube
Sample Preparation:
Centrifuge cup Placement
Cuvelte
Centrifuge Cup

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray SPOTCHEM EZ and is the answer not in the manual?

Arkray SPOTCHEM EZ Specifications

General IconGeneral
BrandArkray
ModelSPOTCHEM EZ
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals