What to do if Arkray Measuring Instruments show 'Test strip feeding trouble'?
- LLisa DixonSep 22, 2025
If a test strip was misfed and fell or moved to an incorrect position, or a test strip was misfed and rebounded, which was detected by the photometric section, or the photometric section detected test strips overflowing from the waste box, or the test strip sensor malfunctioned electrically, or the photometric section malfunctioned electrically, turn OFF the instrument and open the maintenance cover. If there are any obstacles or scattered test strips inside, remove them. Check if the suction ports are damaged. Check whether the feeder is free from adhered urine. If the feeder is dirty, clean it. If the waste box is full of used test strips, empty the box.