Table of Contents

Questions and Answers

  • E
    Emily WilsonAug 8, 2025
    What to do if the printer is not working correctly with Arkray Measuring Instruments?
    • T
      Tammy LopezAug 8, 2025
      If the printer isn't working correctly with your Arkray Measuring Instruments, first, return to the standby screen. Then, ensure the printer is properly attached to the instrument. Clean any dirt from the gold metallic printer terminal on the instrument and the gold four-wire terminal on the printer using a dry, soft cloth or cotton swab.
  • A
    amartinAug 12, 2025
    What to do if Arkray Measuring Instruments shows error E006?
    • B
      Benjamin HarrisAug 13, 2025
      If your Arkray Measuring Instruments device shows error E006, it means the test strip isn't correctly positioned. Return to the standby screen and check that the test strip tip touches the end of the grooves on the holder. Remove any sample or contamination buildup from the grooves.
  • D
    Daniel BrownAug 18, 2025
    What to do if Arkray Measuring Instruments displays error E004?
    • J
      Jason MendezAug 18, 2025
      If your Arkray Measuring Instruments device displays error E004, it means the printer is out of paper. Press the button to return to the standby screen, then load printer paper and press the FEED button.
  • P
    Patrick MarksAug 23, 2025
    What to do if the LED emits too much light on Arkray Measuring Instruments?
    • M
      Marcia MartinezAug 23, 2025
      If the LED emits too much light for testing with your Arkray Measuring Instruments, turn off the power. Clean the white plate (white circle portion) on the test strip holder using a soft cloth dampened with water or alcohol. Then, clean the reading unit.
  • C
    Crystal CohenAug 28, 2025
    What to do if Arkray Measuring Instruments shows error code E007?
    • A
      Ann RussellAug 28, 2025
      If your Arkray Measuring Instruments shows error code E007, it means the test strip type doesn't match what's selected. Check that you've selected the correct test strip type and that the test strip is compatible with the instrument.
  • K
    Kevin MitchellSep 2, 2025
    What to do if Arkray Measuring Instruments displays error E001?
    • F
      Frederick Williams DVMSep 2, 2025
      If your Arkray Measuring Instruments displays error E001, it means the test strip holder isn't moving correctly. Check if the holder is correctly attached and cycle the power. Clean the holder base and guide rails.

Summary

Installation Precautions

Before Use

Measurement Principle

Explains the dual/single wavelength reflectance method used for measurements.

Instrument Setup Procedures

Introduces setup including battery operation and printer attachment.

AC-Power Operation and Printer Detachment

Instructions for AC power connection and detaching the printer.

Basic Operations and Display

Introduces basic operations and explains the instrument's display components.

Operator Panel Functions

Describes the function of each button on the operator panel.

Measurement

Operation Flow Chart

Provides a flowchart of the entire testing procedure.

Cautions in Measurement

General cautions before and during measurement.

Handling Samples

Detailed precautions for handling urine samples.

Handling Test Strips

Guidelines for using and handling test strips.

Test Conditions Setup

Steps to configure test conditions before measurement.

Selecting Test Strip Type

How to select the appropriate test strip type.

Setting Measurement Number

Procedure to set the measurement number for samples.

Setting Patient ID

How to set a patient ID for sample identification.

Normal Measurement Precautions

Precautions before preparing samples and conducting tests.

Preparing Samples and Test Strips

Steps for collecting, preparing urine samples, and handling test strips.

Testing Samples Procedure

Detailed procedure for dipping the test strip and placing it in the instrument.

Dipping Test Strip and Blotting

Specific instructions for dipping the test strip for the correct duration and blotting.

Placing Test Strip and Result Check

How to place the strip and what happens when the test is complete.

Maintenance

Troubleshooting

Arkray PU-4010 Specifications

General IconGeneral
SampleWhole blood, plasma, serum
Measurement itemsGlucose, Lactate
Measurement time (Lactate)60 seconds
Sample volume5 µL
Operating Humidity20 to 80% RH (non-condensing)
Battery LifeApproximately 1000 tests
Operating temperature10 - 35 °C