PU-4010 1-15
1-4-5 Changing the Direction of the Test Strip Feeding
This instrument has been factory-set so that you can place the test strip on the test strip holder
from the right side of the instrument. You can change the direction to the opposite. This
section describes how to change the direction from “the right” to “the left” as an example.
Turn on the power and open the reading
unit cover.
The test strip holder will extend from the
right of the instrument.
Press and hold the button.
The test strip holder will slightly move to the
right until a metallic protrusion appears.
Slide the test strip holder base to the right and
remove it.
Remove the black holder from the holder
base. Reattach the holder to the holder
base so that the white plate comes to the
right side. Be sure to lock the block holder
to the gray base.
Insert the holder base from the right side of
the instrument until it stops.
Change the setting for the direction of the test strip feeding.
See “3-10. Customizing the Instrument” on page 3-11.
The holder base will move to the left.
Test strip
holder
base
Protrusion
1
2
[
6
( )
]
White
plate
3
4
5