PU-4010 1-3
1-1-2 Specifications
Instrument Specifications
Product Compact Urine Analyzer PocketChem UA PU-4010
Configuration Instrument (analyzer) and accessories
Sample Urine
Test strips AUTION Sticks / Uriflet S (9UB)* / AUTION SCREEN
* Except devices with 10V specifications
Measurement items GLU (glucose), PRO (protein), BIL (bilirubin), URO (urobilinogen),
S.G. (specific gravity), BLD (blood), PH (pH), KET (ketones),
NIT (nitrite), LEU (leukocytes), CRE (creatinine) and
ALB (albumin).
Measurement principle Test strip: Dual wavelength reflectance method.
(one wavelength for BLD measurement)
Measurement wavelengths 3 wavelengths LED (565, 635, 760 nm)
Sample supply method Manual dipping
Measurement capacity 50 tests per hour
Test strip reaction time About 60 seconds per test
Measurement types Normal measurement and check measurement
Display Custom LCD that displays information including type of test strip,
measurement number (up to 4 digits), patient ID (up to 13 digits)
and test results.
Operator panel Panel sheet that provides 14 buttons including numeric buttons and
start button.
Memory capacity Test results for 100 samples
Temperature correction Automatically corrected between 10°C to 30°C by internal
temperature sensor
Specific gravity correction Automatically corrected based on pH readings
Chromaturia correction Automatically corrected by color correction test pad
External output Compliant with RS-232C
Operating environment Temperature: 10 to 30ºC; Humidity: 20 to 80% R.H.
(non-condensing)
Measurement environment Temperature: 10 to 30ºC; Humidity: 30 to 60% R.H.
(non-condensing)
*Temperature correction function used
Storage environment Temperature: 1 to 30ºC; Humidity: 20 to 80% R.H.
(non-condensing)
Environment during
transport
Temperature: -10 to 60 ºC; Humidity: 20 to 80% R.H.
(non-condensing)
Sound pressure level Less than 85 dB
Dimensions 124 (W) × 81 (D) × 36 (H) mm
Weight 180 g (without batteries)