PU-4010 2-11
Press the button.
• The 3-second countdown timer will start and three short beeps will sound.
When a long beep is heard, immediately
dip the test strip into the sample. When the
beep stops, take the strip out of the
container.
Blot one side of entire test strip to the
tissue paper (or absorbent paper) to
remove excess urine. Do not blot front or
back of test strip.
4
Dip for
2 seconds
Keep the black
marker dry.
5
IMPORTANT:
• The test strip must be dipped into the
sample for 2 seconds exactly.
Shorter dipping may generate
insufficient color development, while
longer dipping may cause the
reagents on the test strip to leak into
the sample, whereby producing
incorrect results.
• Keep the black marker on the test
strip dry.
6
IMPORTANT: The error code “E008” may appear during countdown in step 4 and
later if room light is too strong for the instrument to read the test strip. In this
case, close the reading unit cover and make sure that the error code
disappears. The timer will continue to count down, so you can continue testing.