PU-4010 2-13
If you suspect the obtained results are inaccurate, use the check strip that came with the
instrument to see if the instrument is working properly. The check strip container contains
two gray strips. Use either one for the test.
Wash the test strip holder.
See “4-1. Daily Cleaning” on page 4-1.
Select the type of the test strip, “C”, for
check measurement.
See “2-4-1. Selecting Test Strip Type” on
page 2-5.
Take a check strip out of the container.
Open the reading unit cover.
Press the button.
Three short beeps will sound, followed by a
long beep.
2-6 Check Measurement
1
NOTE: Unless the holder is washed
before check measurement, the check
strip may be contaminated and lose its
effectiveness.
2
Check strip
3
IMPORTANT: Do not touch the surface
of the check strip. A check strip
contaminated by sebum may cause
incorrect results.
Reading unit cover
4
5