PU-4010 2-14
Before the timer reaches “18SEC, place the
test strip on the test strip holder.
Slide the strip to the left along the grooves of
the holder
as far as it will go.
When the timer reaches “17SEC”, the holder
will move to the left and back again, thus the
instrument will read the test strip.
A short beep will sound and results will print when testing is complete.
Compare results to ranges shown on the
check strip bottle label to see if they fall
within range.
Within range:
Indicates instrument is working properly.
This completes check measurement.
Outside of range:
Indicates either the instrument or the check
strip may not be working properly. Repeat
test with another check strip and go to step 9.
Check the results of the second measurement.
Within range:
Indicates check strip used for the first check measurement may not be working properly.
Do not use it further.
Outside of range:
Indicates instrument may be malfunctioning. Contact your distributor.
Contact here
6
IMPORTANT: An incorrectly positioned
test strip may cause incorrect
results.
7
Qualitative value label
8
Normal check measurement result (example)
* The BLD value will read “0.06 mg/dL” on the
printout.