PU-4010 2-7
2-4-3 Setting Patient ID
If your laboratory or office is managing samples based on patient IDs, you can set the patient
ID for each sample. A patient ID may contain up to 13 digits including numbers (0 to 9) and
hyphens (-).
REFERENCE: The programmed patient ID will be cleared after the completion of a single-
sample measurement. Enter a patient ID each time you measure a new sample.
On the standby screen, press the
button twice.
The last patient ID will appear and the
icon will blink. If no patient ID has been set,
hyphens (-) will appear instead as shown in
the figure.
Press the button.
The cursor ( _ ) will blink.
The last patient ID will be cleared, if one was
set.
Enter the patient ID using the numeric
buttons or hyphen button.
A patient ID may contain up to 13 digits,
including hyphens (-).
REFERENCE: To delete the number one digit at a time, press and release the button. To
clear the entire entry, press and hold the button for one second.
Press the button.
This will confirm your entry and the standby
screen will appear again.
Once the patient ID has been set, the
measurement number and patient ID will
alternately appear on the standby screen.
Go to “2-5. Normal Measurement” on the next page.
1
[
6
( )
]
2
3
Standby screen showing the patient ID