PU-4010 2-10
2-5-3 Testing Samples
Once the sample and test strip have been prepared, dip the test strip into the sample and test
it. Quick, precise handwork is required especially for dipping. Therefore, read this section
thoroughly before testing.
REFERENCE: You can manually stop measurement in progress anytime after pressing the
button in step 4. To do that, press and hold the button for one second or more.
Open the reading unit cover.
Keep one or two pieces of tissue paper (or absorbent paper) at hand.
• This paper is used to remove excess sample from the test strip in later steps.
Hold the test strip. Do not dip it into the sample at this point.
7 cm
(3 inches)
7 cm
(3 inches)
IMPORTANT: For the test strip holder to move freely, leave at least 7 cm (3 inches) of
space to the right and left side of the instrument.
1
2
3