PU-4010 2-12
Before the timer reaches “18SEC”, place the
test strip on the test strip holder.
• Slide the strip to the left along the grooves of
the test strip holder
as far as it will go (see the
figure at below right).
• When the timer reaches “17SEC”, the holder
will move to the left and back again, thus the
instrument will read the test strip.
A short beep will sound when the test is
complete.
Check the result.
• See “2-7. Examples of Results” on page 2-15.
• The result will print from the printer, if attached. Pull the paper toward you and cut it
accordingly.
• The result will also be transmitted to an external device, if connected.
Remove test strip from holder and dispose.
This completes testing.
• To test another sample, return to step 2 on page 2-10.
• If all urine tests has been completed, wash the test strip holder (See “4-1. Daily Cleaning”
on page 4-1).
Set the strip before 18
SEC
.
Contact here.
7
IMPORTANT:
• Be sure to place the test strip before
“18
SEC”. At “17SEC”, the test strip
holder will start moving and you can
no longer place the test strip on the
holder.
• An incorrectly positioned test strip
may cause incorrect results.
8
9
Discard used test strips in accordance with local regulations for
biohazardous waste.
IMPORTANT: Wipe adhered sample off the test strip holder.