6300 Series ComboSource User’s Manual · Page 21
Menu
Description
Factory
Default
TEC Menu » Advanced » User Cal
Edit Cal
Must be set to ‘Yes’ before user calibration settings
can be changed.
No
Reset Cal
Reset the TEC user calibration data to factory
defaults (all slopes to 1, all offsets to 0). Requires
Edit Cal to be set to Yes before this setting will be
displayed.
No
Various
M & B
Settings
Slope (M) and Offset (B) user calibration settings
for each set point and measurement. See the User
Cal - Field Calibration section below for additional
details on using slope/offset calibration.
No
Comm Menu
Baud
This sets the baud rate for the RS-232 serial port.
See the Computer Interfacing Manual which is
included on the CD that accompanied this
product.
9600
Set While
Rmt
To disable changing the set point while in Remote
Mode, set this value to “No”. To enable changing
the set point while in Remote mode, set this Value
to “Yes”.
No
Err While
Rmt
To turn off the display of errors while in remote
mode, set this value to “No”. To display errors
while in remote mode, set this value to “Yes”.
Yes
Terminal
Mode
Terminal mode simply echoes any characters
received over the serial or USB interfaces.
No
Msg Term
This controls the output message termination, and
can be set to CR/LF, CR, LF, or None.
CR/LF
Function Menu
Key A
to
Key D
The Key A, Key B, Key C, and Key D settings
control the action assignment for the four function
keys available on the ComboSource. See Using
the Function Keys below for additional information.
None
System Menu
Config
Allows for restoring the instrument to defaults,
saving configurations, and restoring
configurations. See Saving and Restoring
Configurations below for more information.
No Change